Configuration
I
The 40 germanium detectors are placed at 27
cm from
the target position and cover a total solid angle of 10%
corresponding to a total
absolute photopeak efficiency of 3%
at a gamma-ray energy of 1332
keV.
For the 60Co
source the Compton
suppressed spectra have a mean P/T ratio of 60--65
%.
The average energy resolution
at 1332 keV is better than 2.3
keV and the average relative efficiency
about
82%..
A BGO
multiplicity filter, made of 80 BGO cristals
acts
as a active collimator and measures the total energy and multiplicity of
the
gamma cascade.